Model 14 Oxygen Analyzer
Chapter 8: Error Messages
During failure, the measurement is replaced by dashes “ - - -.”
8-1
CONCENTRATION
TOO HIGH
- - -
STAND-BY
1.99 ppm
- - -
The concentration value exceeds the limits. Check
the current value and the calibration parameters.
Electrode protection
If the measuring range displays “concentration too
high” for more than 2 minutes, the error message
“STAND-BY” appears. From this moment on, the
electrode polarization voltage is switched off for
10 minutes before the voltage is switched on
again.
In STAND-BY the user may switch the
voltage on at any time by pressing ENTER.
The concentration value exceeds the limits. Check
the current value and the calibration parameters.
The sample temperature exceeds the limits. Check
the cable polarity. Check for a short-circuit or an
open circuit.
The current value exceeds the limits. Check the
electrode (electrolyte and membrane).
Error Message
Description
Error messages during a measurement