Broadley Technologies 14 Instrukcja Obsługi Strona 38

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 37
Model 14 Oxygen Analyzer
Chapter 6:
Calibrating the Instrument
6.1. Calibrating the Temperature Sensor
The temperature sensor is located in the dissolved oxygen measurement probe. It is pre-set by the factory, but needs
to be calibrated in the sample, on-site. The temperature sensor must be calibrated before calibrating the sensor’s
slope and zero.
6-1
NOTE
See Chapter 5 for programming commands.
REMARK
Any result (calibration or measurement) is always brought back to the
reference temperature (25˚C, 77˚F). If the sample temperature is different
from the reference temperature, manual or automatic temperature
compensation must be used.
Przeglądanie stron 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 55 56

Komentarze do niniejszej Instrukcji

Brak uwag