5.1.4.6 mA ADJUST Menu
The analog output signals are factory-adjusted (upper limit: 20mA). However, if the upper limit of one of the
outputs, drifts, the span value can be adjusted with the mA ADJUST menu. Connect an ampere meter in series
to the analog output terminals. Adjust the value until the ampere meter displays 20mA.
Disp2 Menu
CONFIGURATION CONFIGURATION
Select
FREQ: 60Hz
FREQ: 50Hz
Enter
Disp2 Menu
Please
wait …
Start up …
Disp2 Menu
Enter
SERVICE
Select
AVERAGE
DISPLAY
CODE
SOFT ISSUE
DEFAULT VAL.
ADJUST mA
CONFIGURATION
BROADLEY J.
5.1.4.7 CONFIGURATION Menu
The operating frequency must be programmed to match the power supply. When the frequency is changed, the
instrument is automatically reset.
ADJUST mA
Select
OUTPUT 1
OUTPUT 2
Enter
ADJUST mA
VALUE: 0000
- +
Enter
SERVICE
Select
AVERAGE
DISPLAY
CODE
SOFT ISSUE
DEFAULT VAL.
ADJUST mA
CONFIGURATION
BROADLEY J.
Note: The value displayed does not correspond to an actual mA value.
Model 14 Oxygen Analyzer
5-16