Model 14 Oxygen Analyzer
3-5
3.11 Programming the Power Supply Frequency
The power supply frequency can be changed if necessary. This change occurs at the initial start-
up and after resetting the instrument. Follow the procedure below.
23.2°CO2
MENU
CALIBRATION
MAINTENANCE
PROGRAMMING
SERVICE
Select
FREQ: 50Hz
CONFIGURATION
Select
Disp2 Menu
Select
SERVICE
AVERAGE
DISPLAY
CODE
SOFT ISSUE
DEFAULT VAL.
ADJUST mA
CONFIGURATION
DISPLAY
Enter
Enter
Enter
Enter
FREQ: 60Hz
FREQ: 60Hz
100
% Sat
DISPLAY
Esc
(press 3 times)